The first procedure to treat
coronary occlusion was a
coronary artery bypass graft (CABG), which uses a section of another vein or artery to bypass the obstructed segment of coronary artery. In 1977,
Andreas Grüntzig introduced
percutaneous transluminal coronary angioplasty (PTCA), also called balloon angioplasty, expanding a
balloon catheter, guided through a peripheral artery, to force expansion of the narrowed segment. As equipment and techniques improved, the use of PTCA rapidly increased, and by the mid-1980s, PTCA and CABG were being performed at equivalent rates. Balloon angioplasty was generally effective and safe, but restenosis was frequent, occurring in about 30–40% of cases, usually within the first year. In about 3% of balloon angioplasty cases, failure of the dilation and acute or threatened closure of the coronary artery (often because of
dissection) prompted emergency CABGs. In 1986, Puel and Sigwart implanted the first coronary stent in a human patient. Several trials in the 1990s showed the superiority of stent placement over balloon angioplasty. Restenosis was reduced because the stent acted as a scaffold to hold open the dilated segment of the artery. Acute closure of the coronary artery (and the requirement for emergency CABG) was reduced, because the stent repaired dissections of the arterial wall. By 1999, stents were used in 84% of
percutaneous coronary interventions (i.e., those done via a catheter, and not by open-chest surgery). As currently used in clinical practice, "drug-eluting" stents refers to metal stents that elute a drug designed to limit the growth of neointimal scar tissue, thus reducing the likelihood of stent
restenosis. The first type of DES to be approved by the
European Medicines Agency (EMA) and the
US Food and Drug Administration (FDA) were sirolimus-eluting stents (SES), releasing
sirolimus, a natural
immunosuppressant drug. SES were shown to reduce the need for repeat procedures and improve the outcomes of patients with coronary artery disease. The sirolimus-eluting Cypher stent received CE mark approval in Europe in 2002, and then underwent a larger trial to demonstrate its safety and effectiveness for the US market. The trial, published in 2003, enrolled 1058 patients with more complex lesions and confirmed the superiority of SES over bare metal stents in terms of angiographic and clinical outcomes. Based on these results, the Cypher stent received FDA approval and was released in the US in 2003. after a series of trials that compared it with a bare metal stent in various settings. The trials showed a significant reduction in target lesion revascularization and major adverse cardiac events with the Taxus stent at 9 and 12 months. Both SES and PES use natural products as the active agents to prevent the recurrence of blockages in the arteries. The initial rapid acceptance of DES led to their peak usage in 2005, accounting for 90% of all stent implantations, but concerns about late stent thrombosis led to a decrease in DES usage in late 2006. Subsequent studies reassured the medical community about their safety, showing that while DES may have a slightly higher risk for very late stent thrombosis, they significantly reduce target vessel revascularization without increasing the incidence of death or myocardial infarction; these reassurances led to a resurgence in DES utilization, although it did not reach the peak usage rates seen in early 2006. The concept of using absorbable (also called biodegradable, bioabsorbable or bioresorbable) materials in stents was first reported in 1878 by Huse who used magnesium wires as ligatures to halt the bleeding in vessels of three patients. Despite extensive search, the full name of this pioneer in the field remains elusive. In 20th century, a resorbable stent tested in humans was developed by the Igaki Medical Planning Company in Japan and was constructed from poly-L-lactic acid (a form of
polylactic acid); they published their initial results in 2000. The German company
Biotronik developed a
magnesium absorbable (bioresorbable) stent and published clinical results in 2007. Despite the initial promise, the first-generation bioresorbable stents, such as the Absorb bioresorbable stent by Abbott, faced significant challenges in their performance. In comparison to current-generation drug-eluting stents, numerous trials revealed that these first-generation bioresorbable stents exhibited poor outcomes. Specifically, they showed high rates of stent thrombosis (cases where an implanted coronary stent caused a thrombotic occlusion), target-lesion myocardial infarction (heart attack occurring at the site of the treated lesion), and target vessel revascularization (the need for further procedures to restore blood flow in the treated artery). In 2017, Abbott pulled its bioabsorbable stent, Absorb, from the European market after negative press regarding the device.
Boston Scientific also announced termination of its Renuvia bioresorbable coronary stent program as studies showed higher risk of serious adverse events. Currently, fully bioresorbable stents do not play a significant role in coronary interventions. While various manufacturers are proposing new stents and continuing their development, it remains uncertain whether they will have a substantial impact, unless there will be more data from their clinical trials. As of now, these stents are not widely utilized in practice. Due to challenges in developing resorbable stents, many manufacturers have focused efforts on targeting or reducing drug release through bioabsorbable-polymer coatings. Boston Scientific's Synergy bioabsorbable polymer stent has been shown potential to reduce the length of
dual antiplatelet therapy post-implantation.
MicroPort's Firehawk target eluting stent has been shown to be non-inferior to traditional drug-eluting stents while using one-third of the amount of equivalent drug. As for the materials used to make a DES, the first DES products available for treating patients were stainless steel alloys composed of iron, nickel, and chromium and were based on existing bare metal stents. Materials explored for use include
magnesium,
polylactic acid,
polycarbonate polymers, and
salicylic acid polymers. For the coating of DES, one to three or more layers of
polymer can be used: a base layer for adhesion, a main layer that holds and elutes (releases) the drug into the arterial wall by contact transfer, and sometimes a top coat to slow down the release of the drug and extend its effect. The first few drug-eluting stents to be licensed used durable coatings. The first generation of coatings appears to have caused immunological reactions at times, and some possibly led to thrombosis. This has driven experimentation and the development of new coating approaches. ==Research directions==