As mentioned above, the success of
CABG surgery may be influenced by the quality of the "conduit" vessel and how it is handled or treated during the harvest and preparation steps prior to grafting. The harvested blood vessel used in coronary artery bypass graft surgery must be free from damage to ensure proper long-term function and good patient outcomes. Conduit quality is a significant factor in long-term patient results. Conduit quality is not always visibly evident when looking at the exterior of the harvested vessel. Damage to the
endothelium, the interior of the vessel, has been shown to increase the likelihood of graft occlusion or blockage. Specifically, damage can be caused during the procedure by: • Extent of thermal injury during the branch division, cutting and sealing • Overhandling of the vessel – during and after harvest • Overdistension of the vessel – when flushing as part of graft preparation • Storage conditions between harvest and graft procedure To preserve and optimize conduit quality, clinical specialists take care to avoid unnecessary thermal injury, overhandling and overdistension, and ensure proper storage conditions for the harvested vessel.
Thermal Injury Comprehensive training and a careful approach are needed when cutting and removing side branches from the main vessel being harvested. As branches are cut, they must be sealed, or cauterized, to prevent bleeding after disconnection from the main vessel. In surgical practice it is common to use a form of heat or electrical energy to
cauterize, or seal tissues in order to stop bleeding. If not properly controlled, the application of heat or energy to seal off the branch vessels can cause unintended thermal injury to the surrounding cells including the
endothelium of the main vessel or conduit. There are two main forms of electrical energy used for cutting and sealing off branch vessels during harvesting: bipolar radio frequency (RF) and direct current (DC).
Bipolar RF Bipolar RF tools pass
alternating electric current (AC) through the tissue – in this case, the branch vessel – located between two electrodes. As the energy passes through the vessel tissue, water molecules within the cells begin to vibrate rapidly, creating heat, resulting in tissue
vaporization and
coagulation. The amount of cutting (vaporization) and sealing (coagulations) is defined by the wave form of the RF energy passed through the tissue.
Direct Current With
direct current devices, current flows to the heating elements contained within a set of "jaws" that clamp down on the branch vessel. The amount of heat generated by the heating elements defines the cutting and sealing. The combination of the heat and pressure results in a single, simultaneous cut-and-seal motion. The distance between the location at which the side branch is cut and the main conduit is especially critical to minimize thermal injury to the main vessel during dissection. On some devices, a mechanical feature is used to physically distance the main conduit from the cautery device.
Overhandling It is important to minimize handling, which may damage or cause strain on the internal endothelial layer of the conduit. Harvester training and experience reduce the likelihood of damage. The risk of damage can also be lessened with techniques such as EVH (rather than bridging), or by using EVH devices that minimize the amount of torque or stretch to the main conduit during harvesting.
Overdistension When a vessel is prepared for use as a bypass graft, a standard syringe is typically used to flush the vessel and check for leaks. Using uncontrolled pressure to flush or clear the vessel can result in damage to the internal cellular lining of the vessel, known as the endothelium. Limiting the maximum pressure that can be applied to the vessel may prevent injury and improve the overall quality and long term patency of the graft. Pressure-limiting syringes are available, and they have been clinically shown to protect against overdistension.
Storage Conditions Studies have indicated that endothelium and
smooth muscle cells are affected by the type of storage solution, and may play a role in long-term vessel
patency. ==References==