Gain in the detection The strength of the difference frequency signal can be larger than the strength of the original signal. The strength of the difference frequency is proportional to the product of the
amplitudes of the LO and original signal electric fields. Thus the larger the LO amplitude, the stronger the difference frequency strength. Hence there is gain in the photon conversion process. The
Poynting vector S of the sum of the LO and signal is proportional to the square of the sum: :S \propto \left[ E_\mathrm{sig}\cos(\omega_\mathrm{sig}t+\varphi_\mathrm{sig}) + E_\mathrm{LO}\cos(\omega_\mathrm{LO}t+\varphi_\mathrm{LO}) \right]^2 = E_\mathrm{sig}^2\cos^2(\omega_\mathrm{sig}t+\varphi_\mathrm{sig}) + E_\mathrm{LO}^2\cos^2(\omega_\mathrm{LO}t+\varphi_\mathrm{LO}) + 2E_\mathrm{sig}E_\mathrm{LO}\cos(\omega_\mathrm{sig}t+\varphi_\mathrm{sig})\cos(\omega_\mathrm{LO}t+\varphi_\mathrm{LO}). A photodetector, like a
photodiode, is much slower than optical frequency (4.75×1014 Hz at 632 nm in wavelength as a red color), so the detector integrates the Poynting vector over a time window much longer than an optical period (2.1×10−15 second for 632 nm in wavelength). In this sense, we can only consider the average of the Poynting vector over the integration window. By using
the product-to-sum trigonometric identity and assuming that the detector is much faster than the difference frequency, the detector output is proportional to the following. (The sum of integrations of terms with \cos at frequencies over a time window much longer than the periods of the frequencies statistically gives zero or a noise in values.) S_\mathrm{avg} = \frac{1}{2}E_\mathrm{sig}^2+\frac{1}{2}E_\mathrm{LO}^2+E_\mathrm{LO}E_\mathrm{sig}\cos((\omega_\mathrm{sig} - \omega_\mathrm{LO})t + \Delta\varphi) where \Delta\varphi = \varphi_\mathrm{sig} - \varphi_\mathrm{LO}. The first two terms are proportional to the average (DC) energy flux absorbed (or, equivalently, the average current in the case of photon counting). The third term is the difference frequency. In many applications the signal is weaker than the LO, thus it can be seen that gain occurs because the energy flux in the difference frequency E_\mathrm{LO}E_\mathrm{sig} is greater than the DC energy flux of the signal \frac{1}{2}E_\mathrm{sig}^2.
Doppler effect Suppose that a signal light in form of E_\mathrm{sig} \cos \left ( \omega_\mathrm{sig} t - k_\mathrm{sig} \int n \left( x \right) x + \varphi_\mathrm{sig} \right ), where k_\mathrm{sig} is the wavenumber in vacuum, is toward a moving target mirror which position is x_m(t). The mirror receives the light at time t_m as E_m \left( t_m \right) = E_\mathrm{sig} \cos \left (\omega_\mathrm{sig} t_m -k_\mathrm{sig} n_{md} x_m(t_m) - k_{\mathrm{sig}} \sum_{i} \int_{l_i} \left[ n_i \left( x \right) - n_{md} \right]dx + \sum_j \Delta \varphi_{j,refl} + \varphi_\mathrm{sig} \right) where n_{md} is the
refractive index of the main wave propagation medium, n_i \left( x \right) is a generally inhomogeneous refractive index of a transmissive optical element
i with its length l_i, and \Delta \varphi_{j,refl} is a phase shift by a reflective optical element
j (
Mirror reflection phase shift). The target mirror immediately (ignoring the
time dilation in the
special relativity) reflects the light, so the reflected light is E_{\mathrm{refl},m}\left ( t_m \right ) = E_\mathrm{sig} \cos \left ( \omega_\mathrm{sig} t_m - k_\mathrm{sig} n_{md} x_m +\varphi_\mathrm{sig}' \right ) where (t_m) in x_m(t_m) is dropped for simplicity, and \varphi_\mathrm{sig}' = - k_{\mathrm{sig}} \sum_{i} \int_{l_i} \left[ n_i \left( x \right) - n_{md} \right]dx + \sum_j \Delta \varphi_{j,refl} + \varphi_\mathrm{sig} with now that the second summation (with the index
j) includes a phase shift from the target mirror reflection. The angular frequency of the reflected light is given by the
time derivative of its phase; \omega_{\mathrm{refl},m} \left( t_m \right ) = \frac{d}{d t_m} \left[ \omega_\mathrm{sig} t_m - k_\mathrm{sig} n_{md} x_m + \varphi_\mathrm{sig}' \right] = \omega_\mathrm{sig} - k_\mathrm{sig} n_{md} v_m = \omega_\mathrm{sig} \left[ 1 - \frac{v_m}{c / n_{md}} \right] where v_m is the mirror velocity at t_m. This is the
Doppler-shifted frequency by the mirror in moving w.r.t the light source. The reflected light is now toward a detector fixed at the location x_d. The detector receives this light at time t = t_m + \frac{x_m - x_d}{c / n_{md,shift}} + \Delta t_{delay} where \Delta t_{delay} = \sum_{p} \frac{\int_{l_p} \left[ n_{p,shift} \left( x \right) - n_{md,shift} \right]dx}{c} is an additional delay by the frequency-shifted light travelling through each transmissive optical element
p till the detector. By assuming constant refractive indices and reflection phase changes over a laser frequency range made by the Doppler shift is assumed., the detected light is \begin{align} E_\mathrm{det} \left( t \right) &= E_\mathrm{sig} \cos \left( \omega_\mathrm{sig} \left[ t - \frac{x_m - x_d}{c / n_{md}} - \Delta t_{delay} \right] - k_\mathrm{sig} n_{md} x_m + \varphi_\mathrm{sig}' \right) \\ &= E_\mathrm{sig} \cos \left( \omega_\mathrm{sig} t - k_\mathrm{sig} n_{md} \left[ x_m - x_d \right] - k_\mathrm{sig} n_{md} x_m +\varphi_\mathrm{sig}'' \right) \\ &= E_\mathrm{sig} \cos \left ( \omega_\mathrm{sig} t - 2k_\mathrm{sig} n_{md} x_m + k_\mathrm{sig} n_{md} x_d + \varphi_\mathrm{sig}'' \right ) \end{align} where \varphi_\mathrm{sig}'' = \varphi_\mathrm{sig}' - \omega_\mathrm{sig} \Delta t_{delay} = - k_{\mathrm{sig}} \sum_{i} \int_{l_i} \left[ n_i \left( x \right) - n_{md} \right]dx + \sum_j \Delta \varphi_{j,refl} + \varphi_\mathrm{sig} - \omega_{\mathrm{sig}}\sum_{p} \frac{\int_{l_p} \left[ n_p \left( x \right) - n_{md} \right]dx}{c} + \sum_q \Delta \varphi_{q,refl}in which \sum_q \Delta \varphi_{q,refl} counts for the phase changes by reflective optics in the way from the target mirror to the detector. The angular frequency of the detected signal light is given by the time derivative of its phase; \omega_\mathrm{det} \left ( t \right ) = \frac{d}{dt} \left[ \omega_\mathrm{sig} t - 2k_\mathrm{sig} n_{md} x_m + k_\mathrm{sig} n_{md} x_d + \varphi_\mathrm{sig}'' \right] = \omega_\mathrm{sig} - 2k_\mathrm{sig} n_{md} v_m = \omega_\mathrm{sig} \left[ 1 - 2 \frac{v_m}{c / n_{md}} \right] because \frac{d x_m}{dt} = \frac{d x_m}{d t_m} \frac{d t_m}{dt} = v_m (\frac{d t_m}{dt} = 1 by ignoring the time dilation). Note that x_m, v_m, and t_m are
retarded quantities (retarded position, velocity, and time) for the light reaching the detector \left ( x, t \right). Then, with a LO light (that travels through fixed optical elements) at the detector at the time t, E_\mathrm{LO} \cos \left ( \omega_\mathrm{LO} t + \varphi_\mathrm{LO,det} \right ), the heterodyne detection signal is proportional to S_\mathrm{avg,hetero}\left ( t \right ) = E_\mathrm{LO}E_\mathrm{sig}\cos(\left[ \omega_\mathrm{sig} - \omega_\mathrm{LO} \right]t + \Delta\varphi) where \Delta \varphi = - 2k_\mathrm{sig} n_{md} x_m + \left[k_\mathrm{sig} n_{md} x_d + \varphi_\mathrm{sig}'' - \varphi_\mathrm{LO,det} \right] = - 2k_\mathrm{sig} n_{md} x_m + \varphi where \varphi = \left[ \ldots \right] is a constant. The time derivate of the phase of the heterodyne signal is \omega_\mathrm{sig} \left[ 1 - 2 \frac{v_m}{c / n_{md}} \right] - \omega_\mathrm{LO}, which turns to measure the mirror velocity v_m (at the retarded time t_m) if the signal and LO frequencies are known, and the integration of the velocity gives the mirror displacement w.r.t a measurement starting time. Note again that the measured mirror velocity (and the mirror displacement) are retarded quantities, at t_m = t - \left[ \frac{x_m - x_d}{c / n_{md}} + \Delta t_\mathrm{delay} \right]. So far, one round travel (positive integer
n = 1) to the moving target mirror by the signal light is considered, giving the Doppler-shifted angular frequency at the detector \omega_\mathrm{det} \left ( t \right ) = \omega_\mathrm{sig} \left[ 1 - 2 \frac{v_m}{c / n_{md}} \right]. For two round travels (
n = 2), the Doppler-shifted light will again go to the target mirror, reflected again toward the detector, so undergo another doppler-shift. For
n round travels (e.g.,
n = 2 for 2-paths interferometer) and that the mirror velocity is about the same during these travels, the detector receives the signal with the frequency of \omega_\mathrm{det} \left ( t \right ) = \omega_\mathrm{sig} \left[ 1 - 2 \frac{v_m}{c / n_{md}} \right]^n. Depending on a purpose of the measurement, an additional method to measure the absolute position of the mirror (that may be with less accuracy) at the heterodyne detection measurement start time may be required to acquire absolute mirror position measurement in a given coordinates system (measured position = initial absolute position + heterodyne-measured relative position change).
Preservation of optical phase The electronically measured signal beam's energy flux, \frac{1}{2} E_\mathrm{sig}^2, is DC and thus erases the phase associated with its optical frequency;
Heterodyne detection allows this phase to be detected. If the optical phase of the signal beam shifts by an angle phi, then the phase of the electronic difference frequency shifts by exactly the same angle phi. More properly, to discuss an optical phase shift one needs to have a common time base reference. Typically the signal beam is derived from the same laser as the LO but shifted by some modulator in frequency. In other cases, the frequency shift may arise from reflection from a moving object. As long as the modulation source maintains a constant offset phase between the LO and signal source, any added optical phase shifts over time arising from external modification of the return signal are added to the phase of the difference frequency and thus are measurable.
Mapping optical frequencies to electronic frequencies allows sensitive measurements As noted above, the difference frequency linewidth can be much smaller than the optical linewidth of the signal and LO signal, provided the two are mutually coherent. Thus small shifts in optical signal center-frequency can be measured: For example, Doppler
lidar systems can discriminate wind velocities with a resolution better than 1 meter per second, which is less than a part in a billion Doppler shift in the optical frequency. Likewise small coherent phase shifts can be measured even for nominally incoherent broadband light, allowing
optical coherence tomography to image micrometer-sized features. Because of this, an
electronic filter can define an effective optical frequency bandpass that is narrower than any realizable wavelength filter operating on the light itself, and thereby enable background light rejection and hence the detection of weak signals.
Noise reduction to shot noise limit As with any small signal amplification, it is most desirable to get gain as close as possible to the initial point of the signal interception: moving the gain ahead of any signal processing reduces the additive contributions of effects like resistor
Johnson–Nyquist noise, or electrical noises in active circuits. In optical heterodyne detection, the mixing-gain happens directly in the physics of the initial photon absorption event, making this ideal. Additionally, to a first approximation, absorption is perfectly quadratic, in contrast to RF detection by a diode non-linearity. One of the virtues of heterodyne detection is that the difference frequency is generally far away
spectrally from the potential noises radiated during the process of generating either the signal or the LO signal, thus the spectral region near the difference frequency may be relatively quiet. Hence, narrow electronic filtering near the difference frequency is highly effective at removing the remaining, generally broadband, noise sources. The primary remaining source of noise is photon shot noise from the nominally constant DC level, which is typically dominated by the Local Oscillator (LO). Since the
shot noise scales as the
amplitude of the LO electric field level, and the heterodyne gain also scales the same way, the ratio of the shot noise to the mixed signal is constant no matter how large the LO. Thus in practice one increases the LO level, until the gain on the signal raises it above all other additive noise sources, leaving only the shot noise. In this limit, the signal to noise ratio is affected by the shot noise of the
signal only (i.e. there is no noise contribution from the powerful LO because it divided out of the ratio). At that point there is no change in the signal to noise as the gain is raised further. (Of course, this is a highly idealized description; practical limits on the LO intensity matter in real detectors and an impure LO might carry some noise at the difference frequency) ==Key problems and their solutions==